Monitoring Radiation Doses during Diagnostic and Therapeutic Neurointerventional Procedures: Multicenter Study for Establishment of Reference Levelsopen accessMonitoring Radiation Doses during Diagnostic and Therapeutic Neurointerventional Procedures: Multicenter Study for Establishment of Reference Levels

Other Titles
Monitoring Radiation Doses during Diagnostic and Therapeutic Neurointerventional Procedures: Multicenter Study for Establishment of Reference Levels
Authors
Yon Kwon Ihn김범수정해웅서상현원유동이영준김동준전평류창우서상일최대섭최시성김상흠변준수노지은송윤선정우상홍노아백상현박정진임수미김정재윤웅
Issue Date
Nov-2021
Publisher
대한신경중재치료의학회
Keywords
Cerebral angiography; Diagnostic reference levels; Radiation monitoring; Intracranial aneurysm; Thrombectomy; Intracranial arteriovenous malformation
Citation
Neurointervention, v.16, no.3, pp 240 - 251
Pages
12
Indexed
SCOPUS
KCICANDI
Journal Title
Neurointervention
Volume
16
Number
3
Start Page
240
End Page
251
URI
https://scholarworks.gnu.ac.kr/handle/sw.gnu/4876
DOI
10.5469/neuroint.2021.00437
ISSN
2093-9043
2233-6273
Abstract
Purpose: To assess patient radiation doses during diagnostic and therapeutic neurointerventional procedures from multiple centers and propose dose reference level (RL). Materials and Methods: Consecutive neurointerventional procedures, performed in 22 hospitals from December 2020 to June 2021, were retrospectively studied. We collected data from a sample of 429 diagnostic and 731 therapeutic procedures. Parameters including dose-area product (DAP), cumulative air kerma (CAK), fluoroscopic time (FT), and total number of image frames (NI) were obtained. RL were calculated as the 3rd quartiles of the distribution. Results: Analysis of 1160 procedures from 22 hospitals confirmed the large variability in patient dose for similar procedures. RLs in terms of DAP, CAK, FT, and NI were 101.6 Gy·cm2, 711.3 mGy, 13.3 minutes, and 637 frames for cerebral angiography, 199.9 Gy·cm2, 3,458.7 mGy, 57.3 minutes, and 1,000 frames for aneurysm coiling, 225.1 Gy·cm2, 1,590 mGy, 44.7 minutes, and 800 frames for stroke thrombolysis, 412.3 Gy·cm2, 4,447.8 mGy, 99.3 minutes, and 1,621.3 frames for arteriovenous malformation (AVM) embolization, respectively. For all procedures, the results were comparable to most of those already published. Statistical analysis showed male and presence of procedural complications were significant factors in aneurysmal coiling. Male, number of passages, and procedural combined technique were significant factors in stroke thrombolysis. In AVM embolization, a significantly higher radiation dose was found in the definitive endovascular cure group. Conclusion: Various RLs introduced in this study promote the optimization of patient doses in diagnostic and therapeutic interventional neuroradiology procedures. Proposed 3rd quartile DAP (Gy·cm2) values were 101.6 for diagnostic cerebral angiography, 199.9 for aneurysm coiling, 225.1 for stroke thrombolysis, and 412.3 for AVM embolization. Continual evolution of practices and technologies requires regular updates of RLs.
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medicine > Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Researcher Choi, Dae Seob photo

Choi, Dae Seob
의과대학 (의학과)
Read more

Altmetrics

Total Views & Downloads

BROWSE