Design and manufacture of calf-link with knee joint torque sensor for a tendon-driven walking assistant robot

Citations

SCOPUS

1

초록

In this paper, the design and fabrication of the calf-link with knee joint torque sensor of a tandem-driven walking-assist robot is described. Tendon-driven walking-assist robots should be designed and constructed with a wire wheel and a torque sensor, as one body to reduce the weight of the calf link. The torque sensor consists of four plate sensing parts crossed 90° around the wire wheel. Structural analysis was performed to determine the size of the torque sensor sensing part, and a torque sensor was built by attaching a strain gauge to the sensing part. As a result of the characteristics test, the reproducibility error and the nonlinearity error of the manufactured torque sensor were less than 0.03% and 0.04%, respectively. As a result of the calibration, the reproducibility error and the nonlinearity error were less than 0.08%, respectively. Thus, it is considered that the knee joint torque sensor of the calf link can be attached to the tandem-driven walking-assist robot. Copyright ? The Korean Society for Precision Engineering.

키워드

Non-linearity error; Rated output; Repeatability error; Strain-gage; Torque sensor; Walking-assist robot
제목
Design and manufacture of calf-link with knee joint torque sensor for a tendon-driven walking assistant robot
저자
An, J.-H.; Kim, G.S.
DOI
10.7736/KSPE.2019.36.11.1009
발행일
2019
유형
Article
저널명
한국정밀공학회지
권
36
호
11
페이지
1009 ~ 1015