Design of a four-axis force/moment sensor for measuring the applied force to wrist

Citations

SCOPUS

2

초록

Patients have the paralysis of their wrists, and can't use of their wrists freely. But their wrists can be recovered by wristbending rehabilitation exercise. Professional rehabilitation therapeutists exercise the wrists of patients in hospital. But the wrists of patients have not exercised enough for the rehabilitation, because the therapeutists are much less than patients in number. Therefore, the wrist rehabilitation robot should be developed, and it have to measure the applied force to the patients' wrists for their safety. In this paper, the four-axis force/moment sensor was designed for the wrist rehabilitation robot. As a test results, the interference error of the four-axis force/moment sensor was less than 0.91%. It is thought that the sensor can be used to measure the applied force to the patients' wrists. ? ICROS 2013.

키워드

Four-axis force/moment sensorInterference errorRated outputWrist rehabilitation robotWrist-bending exercise
제목
Design of a four-axis force/moment sensor for measuring the applied force to wrist
저자
Hong, T.-K.Kim, G.-S.
DOI
10.5302/J.ICROS.2013.13.1924
발행일
2013
유형
Article
저널명
제어.로봇.시스템학회 논문지
19
11
페이지
1011 ~ 1016