상세 보기
중증뇌졸중환자의 발목재활로봇을 위한 힘센서 설계
- 김한솔;
- 김갑순
초록
This paper describes the design and fabrication of a two-axis force/torque sensor and an one-axis force sensor with parallel plate beams(PPSs) for measuring forces and torque in an ankle rehabilitation exercise using by a lower rehabilitation robot. The two-axis force/torque sensor is composed of a Fy force sensor and Tz torque sensor and the force sensor detects x direction force. The two-axis force/torque sensor and one-axis force sensor were designed using by FEM(Finite Element Method), and manufactured using strain-gages. The characteristics experiment of the two-axis force/torque sensor and one-axis force sensor were carried out respectively. As a test results, the interference error of the two-axis force/torque sensor was less than 1.56%, the repeatability error and the non-linearity of the two-axis force/torque sensor were less than 0.03% respectively, and the repeatability error and the non-linearity of the one-axis force sensor were less than 0.03% and 0.02% respectively.
키워드
- 제목
- 중증뇌졸중환자의 발목재활로봇을 위한 힘센서 설계
- 제목 (타언어)
- Design of Force Sensors for the Ankle Rehabilitation Robot of Severe Stroke Patients
- 저자
- 김한솔; 김갑순
- 발행일
- 2016-03
- 저널명
- 센서학회지
- 권
- 25
- 호
- 2
- 페이지
- 148 ~ 154