중증뇌졸중환자의 발목재활로봇을 위한 힘센서 설계

Design of Force Sensors for the Ankle Rehabilitation Robot of Severe Stroke Patients

초록

This paper describes the design and fabrication of a two-axis force/torque sensor and an one-axis force sensor with parallel plate beams(PPSs) for measuring forces and torque in an ankle rehabilitation exercise using by a lower rehabilitation robot. The two-axis force/torque sensor is composed of a Fy force sensor and Tz torque sensor and the force sensor detects x direction force. The two-axis force/torque sensor and one-axis force sensor were designed using by FEM(Finite Element Method), and manufactured using strain-gages. The characteristics experiment of the two-axis force/torque sensor and one-axis force sensor were carried out respectively. As a test results, the interference error of the two-axis force/torque sensor was less than 1.56%, the repeatability error and the non-linearity of the two-axis force/torque sensor were less than 0.03% respectively, and the repeatability error and the non-linearity of the one-axis force sensor were less than 0.03% and 0.02% respectively.

키워드

Two-axis force sensorForce sensorRated outputInterference errorAnkle force
제목
중증뇌졸중환자의 발목재활로봇을 위한 힘센서 설계
제목 (타언어)
Design of Force Sensors for the Ankle Rehabilitation Robot of Severe Stroke Patients
저자
김한솔김갑순
DOI
10.5369/JSST.2016.25.2.148
발행일
2016-03
저널명
센서학회지
25
2
페이지
148 ~ 154