상세 보기
Design of force sensors of rehabilitation robot for knee rehabilitation
- Jung, J.-H.;
- Kim, G.-S.
SCOPUS
5초록
This paper describes the design and fabrication of a one-axis force sensor and two-axis force sensors with parallel plate beams (PPBs) for measuring forces in a knee rehabilitation exercise using a lower rehabilitation robot. The two-axis force sensors are composed of a Fx force sensor and a Fz force sensor, and they detect the x and z directional forces, respectively. The two-axis force sensors and one-axis force sensor were designed using the Finite Element Method (FEM), and manufactured using strain-gages. The characteristics experiment of the two-axis force sensors and one-axis force sensor were performed. The test results indicated that the interference error of the two-axis force sensor was less than 1.17%, the repeatability error and the non-linearity of the one-axis force sensor were less than 0.03%, and the repeatability error and the non-linearity of the two-axis force sensors were less than 0.03%. ? ICROS 2017.
키워드
- 제목
- Design of force sensors of rehabilitation robot for knee rehabilitation
- 저자
- Jung, J.-H.; Kim, G.-S.
- 발행일
- 2017
- 유형
- Article
- 저널명
- 제어.로봇.시스템학회 논문지
- 권
- 23
- 호
- 1
- 페이지
- 46 ~ 52
- 언어
- KOR
- 출판사
- Institute of Control, Robotics and Systems
- 발행국가
- 대한민국
- 분량
- 7 페이지
- ISSN
- P 1976-5622