다리 환자를 위한 보행보조로봇의 종아리 링크 3축 힘센서 개발

Development of Calf Link Force Sensors of Walking Assist Robot for Leg Patients

초록

This paper describes the design and fabrication of a three-axis force sensor with parallel plate beams (PPSs) for measuring the calf force while a patient with a walking assist robot is walking. Current walking assist robots can’t measure the weight of the patient’s legand the robot’s leg which required for robot control. So, the three-axis force sensor in the calf link is designed and manufactured, it iscomposed of a Fx force sensor, a Fy force sensor and a Fz force sensor. The three-axis force sensor was designed using by FEM(FiniteElement Method), and fabricated using strain-gages. The characteristics experiment of the three-axis force sensor was carried out respec-tively. The test results indicated that the repeatability error and the non-linearity error of three-axis force sensor was less than 0.04% respectively. Therefore, the fabricated three-axis force sensor in the calf link can be used to measure the patient’s calf force in the walking assist robot.

키워드

Gas Strain-gageThree-axis force sensorForce sensorRated outputRepeatability errorNon-linearity error
제목
다리 환자를 위한 보행보조로봇의 종아리 링크 3축 힘센서 개발
제목 (타언어)
Development of Calf Link Force Sensors of Walking Assist Robot for Leg Patients