Initial Experience with Left Bundle Branch Area Pacing with Conventional Stylet-Driven Extendable Screw-In Leads and New Pre-Shaped Delivery Sheathsopen access

Authors
Byeon, KyeongminKim, Hye ReePark, Seung-JungPark, Young JunChoi, Ji-HoonKim, Ju YounPark, Kyoung-MinOn, Young KeunKim, June Soo
Issue Date
May-2022
Publisher
MDPI AG
Keywords
left bundle branch area pacing; pacemaker; stylet-driven extendable screw lead; efficacy; safety
Citation
Journal of Clinical Medicine, v.11, no.9
Indexed
SCIE
SCOPUS
Journal Title
Journal of Clinical Medicine
Volume
11
Number
9
URI
https://scholarworks.gnu.ac.kr/handle/sw.gnu/71656
DOI
10.3390/jcm11092483
ISSN
2077-0383
2077-0383
Abstract
Until recently, left bundle branch area pacing (LBBAp) has mostly been performed using lumen-less fixed screw leads. There are limited data on LBBAp with conventional style-driven extendable screw-in (SDES) leads, particularly data performed by operators with no previous experience with LBBAp procedures. In total, 42 consecutive patients undergoing LBBAp using SDES leads and newly designed delivery sheaths (LBBAp group) were compared with those treated with conventional right ventricular pacing (RVp) for atrioventricular block (RVp group, n = 84) using propensity score matching (1:2 ratio). The LBBAp was successful in 83% (35/42) of patients, with satisfactory pacing thresholds (0.8 +/- 0.2 V at 0.4 ms). In the LBBAp group, the mean paced-QRS duration obtained during RV apical pacing (173 +/- 18 ms) was significantly reduced by LBBAp (116 +/- 14 ms, p < 0.001). Compared with the RVp group, the LBBAp group showed more physiological pacing, suggested by a much narrower paced-QRS duration (116 +/- 14 vs. 151 +/- 21 ms, p < 0.001). The pacing threshold was comparable in both groups. The LBBAp group revealed stable pacing thresholds for 6.8 +/- 4.8 months post-implant and no serious complications including lead dislodgement or septal perforation. The novel approach of LBBAp using SDES leads and the new dedicated pre-shaped delivery sheaths was effectively and safely performed, even by inexperienced operators with LBBAp procedures.