Spurious Elevation of Glucose Concentration during Administration of High Dose of Ascorbic Acid in a Patient with Type 2 Diabetes on Hemodialysis

  • Kim, Soo Kyoung
  • Hahm, Jong Ryeal
  • Kim, Ho-Su
  • Kim, Sungsu
  • Jung, Tae Sik
  • ... Kim, Deok Ryong
  • 외 2명
Citations

WEB OF SCIENCE

12
Citations

SCOPUS

15

초록

We describe herein a case of life-threatening hypoglycemia due to spurious elevation of glucose concentration during the administration of ascorbic acid in a type 2 diabetic patient. A 31-year-old female was admitted for proliferative diabetic retinopathy treatment and prescribed high dose ascorbic acid. During hospitalization, she suddenly lost her consciousness and her glucose concentration was 291 mg/dL, measured using self-monitoring blood glucose (SMBG) device, while venous blood glucose concentration was 12 mg/dL. After intravenous injection of 50% glucose solution, the patient became alert. We reasoned that glucose measurement by SMBG device was interfered by ascorbic acid. Physicians should be aware of this interference; high dose ascorbic acid may cause spurious elevation of glucose concentration when measuring with SMBG devices.

키워드

Self-monitoring of blood glucoseascorbic acidhypoglycemiaBLOOD-GLUCOSEPERITONEAL-DIALYSISTEST STRIPSICODEXTRIN
제목
Spurious Elevation of Glucose Concentration during Administration of High Dose of Ascorbic Acid in a Patient with Type 2 Diabetes on Hemodialysis
저자
Kim, Soo KyoungHahm, Jong RyealKim, Ho-SuKim, SungsuJung, Tae SikJung, Jung HwaLee, Hye RyunKim, Deok Ryong
DOI
10.3349/ymj.2013.54.5.1289
발행일
2013-09
유형
Article
저널명
Yonsei Medical Journal
54
5
페이지