상세 보기
초록
Breakthrough analysis has widely been explored for the dynamic separation of gaseous mixtures in porous materials. In general, breakthrough experiments measure the components of a flowing gas when a gaseous mixture is injected into a column filled with an adsorbent material. In this paper, we report on the design and fabrication of a breakthrough curve measurement device to study the dynamic adsorptive separation of hydrogen isotopologues in porous materials. Using the designed system, an experiment was conducted involving a 1:1 mixture of hydrogen and deuterium passed through a column filled with zeolite 13X (1 g). At room temperature, both hydrogen and deuterium were adsorbed in negligible amounts; however, at a temperature of 77 K, deuterium was preferentially adsorbed over hydrogen. The selectivity was different from that in the existing literature due to the different sample shapes, measurement methods, and column structures, but was at a similar level to that of cryogenic distillation (1.5).
키워드
- 제목
- 수소 동위원소 분리를 위한 77 K 극저온 파과 곡선 측정 시스템 제작
- 제목 (타언어)
- Investigation of Cryogenic Breakthrough Curve Measurement System at 77 K for Hydrogen Isotopologue Separation
- 저자
- Kim, Suhwan; Oh, Hyunchul
- 발행일
- 2022-01
- 유형
- Article
- 저널명
- 한국재료학회지
- 권
- 32
- 호
- 1
- 페이지
- 36 ~ 43