This paper reports the results of a validation process for particle analysis of environmental samples for nuclear safeguards using small geometry-secondary ion mass spectrometry. An analytical method was established for particle recovery, particle search, and uranium isotope analysis. Test samples containing uranium and sediment materials were analyzed using the method, and two groups of particles containing low-enriched uranium with U-235 contents in proximity were resolved. The method was validated for environmental analysis using a set of field samples. The results were consistent with previously reported values. The validated method will contribute to the pursuit of global nuclear non-proliferation.

키워드

SafeguardsEnvironmental sample analysisUranium particleUranium isotopesSecondary ion mass spectrometryMethod validationINDIVIDUAL PARTICLESSIMS
제목
Method validation for the analysis of uranium particles in environmental samples using secondary ion mass spectrometry for safeguards purpose
저자
Lee, Na-RiKim, Tae HeeLee, Chi-GyuPark, Jinkyu
DOI
10.1007/s10967-022-08629-y
발행일
2022-12
유형
Article
저널명
Journal of Radioanalytical and Nuclear Chemistry
331
12
페이지
5605 ~ 5612