Midterm Results of Large Diameter Biolox Forte Ceramic Head on Delta Ceramic Liner Articulation in Total Hip Arthroplasty
- Kang, Bun-Jung;
- Ha, Yong-Chan;
- Hwang, Sun-Chul;
- Lee, Young-Kyun;
- Koo, Kyung-Hoi
Citations
WEB OF SCIENCE
12Citations
SCOPUS
15초록
Seventy-nine patients (94 hips), who underwent cementless alumina-on-alumina total hip arthroplasty (THA) with the use of a 36-mm delta ceramic liner-on-alumina ceramic femoral head, were followed for an average of 6.5 years (range, 5-7.7 years). All acetabular and femoral components were bone-ingrown and neither pelvic nor femoral osteolysis was identified until the latest follow-up. The survivorship with revision surgery as an endpoint was 97.9% (95% confidence interval = 100%-95%). Ceramic related complications such as fracture and squeaking did not occur in any patient. The mid-term results of cementless THA with this type of ceramic articulation are encouraging. (C) 2014 Elsevier Inc. All rights reserved.
키워드
total hip arthroplasty; cementless; large diameter femoral head; delta liner; FOLLOW-UP; ACETABULAR COMPONENT; FEMORAL COMPONENTS; BEARING; REPLACEMENT; FIXATION; FRACTURE; MINIMUM; FAILURE; WEAR
- 제목
- Midterm Results of Large Diameter Biolox Forte Ceramic Head on Delta Ceramic Liner Articulation in Total Hip Arthroplasty
- 저자
- Kang, Bun-Jung; Ha, Yong-Chan; Hwang, Sun-Chul; Lee, Young-Kyun; Koo, Kyung-Hoi
- 발행일
- 2014-12
- 유형
- Article
- 권
- 29
- 호
- 12
- 페이지
- 2412 ~ 2414
- 언어
- ENG
- 출판사
- CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
- 발행국가
- 미국
- 분량
- 3 페이지
- ISSN
- E 1532-8406