Cited 0 time in
Biomechanical finite element analysis of a femoral neck system fixation construct for femur neck fractures and clinical implications
| DC Field | Value | Language |
|---|---|---|
| dc.contributor.author | Hoon-Sang Sohn | - |
| dc.contributor.author | Se-Lin Jeong | - |
| dc.contributor.author | Gu-Hee Jung | - |
| dc.date.accessioned | 2025-08-06T07:30:09Z | - |
| dc.date.available | 2025-08-06T07:30:09Z | - |
| dc.date.issued | 2025-07 | - |
| dc.identifier.issn | 3058-6267 | - |
| dc.identifier.issn | 3058-6275 | - |
| dc.identifier.uri | https://scholarworks.gnu.ac.kr/handle/sw.gnu/79642 | - |
| dc.description.abstract | Background: This study assessed the structural/mechanical stability of fixation constructs with a femoral neck system (FNS) via finite element analysis after simulating femoral neck fractures and explored the clinical implications. Methods: We simulated subcapital, transcervical, basicervical, and vertical fracture models using a right femur (SAWBONES) and imported the implant model of FNS to Ansys (Ansys 19.0, Ansys Inc.) to place the implant in the optimal position. The distal end of the femur model was completely fixed and was abducted 7°. The force vector was set laterally at an angle of 3° and posteriorly at an angle of 15° in the vertical ground. The analysis was conducted using Ansys software with the von Mises stress (VMS) in megapascals (MPa). Results: The maximum VMS of the fracture site was 67.01 MPa for a subcapital, 68.56 MPa for a transcervical, 344.54 MPa for a basicervical, and 130.59 MPa for a vertical model. The maximum VMS of FNS was 840.34 MPa for a subcapital, 637.37 MPa for a transcervical, 464.07 MPa for a basicervical, and 421.01 MPa for a vertical model. The stress distribution of basicervical and vertical fractures differed significantly, and the basicervical fracture had higher VMS at the bone, implant, and fracture sites. Conclusions: FNS fixation should be performed with consideration the osseous anchorage in the femoral head, and this technique might be appropriate for vertical fractures. Regarding the VMS at the fracture site, FNS might be applied cautiously only to basicervical fractures with anatomical reduction without a gap or comminution. Level of evidence: IV. | - |
| dc.format.extent | 10 | - |
| dc.language | 영어 | - |
| dc.language.iso | ENG | - |
| dc.publisher | 대한골절학회 | - |
| dc.title | Biomechanical finite element analysis of a femoral neck system fixation construct for femur neck fractures and clinical implications | - |
| dc.type | Article | - |
| dc.publisher.location | 대한민국 | - |
| dc.identifier.doi | 10.12671/jmt.2025.00108 | - |
| dc.identifier.bibliographicCitation | Journal of Musculoskeletal Trauma, v.38, no.3, pp 133 - 142 | - |
| dc.citation.title | Journal of Musculoskeletal Trauma | - |
| dc.citation.volume | 38 | - |
| dc.citation.number | 3 | - |
| dc.citation.startPage | 133 | - |
| dc.citation.endPage | 142 | - |
| dc.type.docType | Y | - |
| dc.identifier.kciid | ART003227520 | - |
| dc.description.isOpenAccess | N | - |
| dc.description.journalRegisteredClass | kci | - |
| dc.subject.keywordAuthor | Proximal femoral fractures | - |
| dc.subject.keywordAuthor | Fracture fixation | - |
| dc.subject.keywordAuthor | Finite element analysis | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
Gyeongsang National University Central Library, 501, Jinju-daero, Jinju-si, Gyeongsangnam-do, 52828, Republic of Korea+82-55-772-0532
COPYRIGHT 2022 GYEONGSANG NATIONAL UNIVERSITY LIBRARY. ALL RIGHTS RESERVED.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.