Cited 30 time in
PVC membrane-based portable ion analyzer for hydroponic and water monitoring
| DC Field | Value | Language |
|---|---|---|
| dc.contributor.author | Kim, Hak-Jin | - |
| dc.contributor.author | Kim, Dong-Wook | - |
| dc.contributor.author | Kim, Won Kyung | - |
| dc.contributor.author | Cho, Woo-Jae | - |
| dc.contributor.author | Kang, Chang Ik | - |
| dc.date.accessioned | 2025-03-20T07:00:13Z | - |
| dc.date.available | 2025-03-20T07:00:13Z | - |
| dc.date.issued | 2017-08 | - |
| dc.identifier.issn | 0168-1699 | - |
| dc.identifier.issn | 1872-7107 | - |
| dc.identifier.uri | https://scholarworks.gnu.ac.kr/handle/sw.gnu/77480 | - |
| dc.description.abstract | Rapid on-site measurement of chemical ions in hydroponic and water solutions would allow efficient management of nutrients used for plant production as well as early monitoring of chemical pollutants. This paper reports the development of an embedded portable analyzer incorporated with a sensor array of polyvinyl chloride (PVC) membrane-based ion-selective electrodes (ISEs) to directly measure the concentrations of NO3, K, and H ions in hydroponic solutions and water. The developed ion analyzer consisted of an AVR microcontroller, five channels of sensor inputs, a 16-bit analog-to-digital converter, a 7-inch LCD touch panel, and SD memory card-based data storage. The use of a two-point normalization method consisting of sensitivity adjustment followed by offset compensation was effective in minimizing signal drifts resulting from a series of sample measurements while reducing variability in response among multiple ISEs during replicate measurements in practical manners. The sensitivity and predictive capability of the PVC membrane-based H ISEs, in conjunction with a three-point calibration method, were satisfactory, showing sub-Nernstian slopes of 51.9-56.1 mV/decade over a pH range of 4-10 (R-2 > 0.96) and providing results in close agreement with the results of a conventional pH meter (a nearly 1:1 regression slope and a y-intercept near 0). A response time of 50 s and a wide sensitivity range of 11884 mg L-1 measured with the prototype portable ion analyzer equipped with PVC membrane-based NO3 and K ISEs, each with a lifetime of 60 days, enabled direct analysis of hydroponic and water samples without the need to dilute samples. Strong linear relationships between the prototype ion analyzer and the standard instrument methods (R-2 > 0.97, slopes ranging from 0.89 to 1.24) were exhibited. (C) 2017 Elsevier B.V. All rights reserved. | - |
| dc.format.extent | 12 | - |
| dc.language | 영어 | - |
| dc.language.iso | ENG | - |
| dc.publisher | Elsevier BV | - |
| dc.title | PVC membrane-based portable ion analyzer for hydroponic and water monitoring | - |
| dc.type | Article | - |
| dc.publisher.location | 영국 | - |
| dc.identifier.doi | 10.1016/j.compag.2017.06.015 | - |
| dc.identifier.scopusid | 2-s2.0-85021163432 | - |
| dc.identifier.wosid | 000407182700035 | - |
| dc.identifier.bibliographicCitation | Computers and Electronics in Agriculture, v.140, pp 374 - 385 | - |
| dc.citation.title | Computers and Electronics in Agriculture | - |
| dc.citation.volume | 140 | - |
| dc.citation.startPage | 374 | - |
| dc.citation.endPage | 385 | - |
| dc.type.docType | Article | - |
| dc.description.isOpenAccess | N | - |
| dc.description.journalRegisteredClass | scie | - |
| dc.description.journalRegisteredClass | scopus | - |
| dc.relation.journalResearchArea | Agriculture | - |
| dc.relation.journalResearchArea | Computer Science | - |
| dc.relation.journalWebOfScienceCategory | Agriculture, Multidisciplinary | - |
| dc.relation.journalWebOfScienceCategory | Computer Science, Interdisciplinary Applications | - |
| dc.subject.keywordPlus | COMPUTER-CONTROLLED SYSTEM | - |
| dc.subject.keywordPlus | MANAGEMENT | - |
| dc.subject.keywordPlus | GREENHOUSE | - |
| dc.subject.keywordPlus | PH | - |
| dc.subject.keywordPlus | ELECTRODE | - |
| dc.subject.keywordPlus | ARRAY | - |
| dc.subject.keywordAuthor | Ion-selective electrode | - |
| dc.subject.keywordAuthor | Two-point normalization | - |
| dc.subject.keywordAuthor | Nitrate | - |
| dc.subject.keywordAuthor | pH | - |
| dc.subject.keywordAuthor | Potassium | - |
Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.
Gyeongsang National University Central Library, 501, Jinju-daero, Jinju-si, Gyeongsangnam-do, 52828, Republic of Korea+82-55-772-0532
COPYRIGHT 2022 GYEONGSANG NATIONAL UNIVERSITY LIBRARY. ALL RIGHTS RESERVED.
Certain data included herein are derived from the © Web of Science of Clarivate Analytics. All rights reserved.
You may not copy or re-distribute this material in whole or in part without the prior written consent of Clarivate Analytics.