dc.description.abstractWe investigated the capability of a trained deep learning (DL) model with a convolutional neural network (CNN) in a different scanning environment in terms of ameliorating the quality of synthetic fluid-attenuated inversion recovery (FLAIR) images. The acquired data of 319 patients obtained from the retrospective review were used as test sets for the already trained DL model to correct the synthetic FLAIR images. Quantitative analyses were performed for native synthetic FLAIR and DL-FLAIR images against conventional FLAIR images. Two neuroradiologists assessed the quality and artifact degree of the native synthetic FLAIR and DL-FLAIR images. The quantitative parameters showed significant improvement on DL-FLAIR in all individual tissue segments and total intracranial tissues than on the native synthetic FLAIR (p < 0.0001). DL-FLAIR images showed improved image quality with fewer artifacts than the native synthetic FLAIR images (p < 0.0001). There was no significant difference in the preservation of the periventricular white matter hyperintensities and lesion conspicuity between the two FLAIR image sets (p = 0.217). The quality of synthetic FLAIR images was improved through artifact correction using the trained DL model on a different scan environment. DL-based correction can be a promising solution for ameliorating the quality of synthetic FLAIR images to broaden the clinical use of synthetic magnetic resonance imaging (MRI).-
dc.language영어-
dc.language.isoENG-
dc.publisherMDPI-
dc.titleValidation of Deep Learning-Based Artifact Correction on Synthetic FLAIR Images in a Different Scanning Environment-
dc.typeArticle-
dc.publisher.location스위스-
dc.identifier.doi10.3390/jcm9020364-
dc.identifier.scopusid2-s2.0-85085276226-
dc.identifier.wosid000518823000076-
dc.identifier.bibliographicCitationJOURNAL OF CLINICAL MEDICINE, v.9, no.2-
dc.citation.titleJOURNAL OF CLINICAL MEDICINE-
dc.citation.volume9-
dc.citation.number2-
dc.type.docTypeArticle-
dc.description.isOpenAccessY-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.relation.journalResearchAreaGeneral & Internal Medicine-
dc.relation.journalWebOfScienceCategoryMedicine, General & Internal-
dc.subject.keywordPlusMULTIPLE-SCLEROSIS-
dc.subject.keywordPlusBRAIN-
dc.subject.keywordPlusMRI-
dc.subject.keywordAuthorneural networks (computer)-
dc.subject.keywordAuthordeep learning-
dc.subject.keywordAuthorimage enhancement-
dc.subject.keywordAuthormagnetic resonance imaging-
dc.subject.keywordAuthorimage interpretation-
dc.subject.keywordAuthorcomputer-assisted-
Files in This Item
There are no files associated with this item.
Appears in
Collections
College of Medicine > Department of Medicine > Journal Articles

qrcode

Items in ScholarWorks are protected by copyright, with all rights reserved, unless otherwise indicated.

Altmetrics

Total Views & Downloads

BROWSE